RPR-16
Lead thickness from TVL
x = TVL log10(1/B). Lead (or concrete) needed for transmission B.
Governing equation
where
- TVL
- Tenth-value layer (mm)
- B
- Required transmission (—)
- x
- Shield thickness (mm)
- n
- Number of TVLs (—)
Lecture brief
Historical brief
Inverse-square, half-value layer, ICRP weighting and ALARA are the protection craft since the 1920s commissions. The lab computes transmission, equivalent dose and a shielding snapshot. This sheet (RPR-16 — Lead thickness from TVL) is the form associated with Lead thickness. Working symbols: , , . First TVL is larger than equilibrium TVL. Diagnostic lead: ~0.3–0.5 mm HVL at 80 kV; linac concrete: tens of cm.
Purpose
Live realistic example
In symbols
Calculator
Inputs
Outputs
- Shield thickness x24.00 mm
- Number of TVLs n2.00 —
Watch on YouTube
Free library
Full libraryFree PDF / open book
- Anatomy and Physiology 2eOpenStax · CC BY · Free PDF / open book
- Radiation Oncology Physics (Podgorsak)IAEA STI/PUB/1196 · Free PDF / open book
- Diagnostic Radiology PhysicsIAEA · Free PDF / open book
- Nuclear Medicine PhysicsIAEA · Free PDF / open book
YouTube channels
Narration of this film
A required B and a stack of tenth-value sheets.
First TVL is larger than equilibrium TVL. Diagnostic lead: ~0.3–0.5 mm HVL at 80 kV; linac concrete: tens of cm.