SAD versus SSD conversion

D_SAD / D_SSD = ((SSD+d)/SAD)². Inverse-square between the two setups.

Reading speed
GeometrySAD/SSD

Governing equation

DSADDSSD=(SSD+dSAD)2\dfrac{D_{\mathrm{SAD}}}{D_{\mathrm{SSD}}}=\left(\dfrac{\mathrm{SSD}+d}{\mathrm{SAD}}\right)^{2}

where

D_{SSD}
SSD dose (Gy)
SSD
SSD (cm)
d
Depth (cm)
SAD
SAD (cm)
D_{SAD}
SAD dose (Gy)

Lecture brief

Historical brief

The linear-quadratic model, BED, tissue-maximum ratio and Clarkson scatter turned radiotherapy into fraction arithmetic. MU calculation on these sheets is that clinical closed form. This sheet (RTH-15 — SAD versus SSD conversion) is the form associated with SAD/SSD. Working symbols: DSSDD_{SSD}, SSDSSD, dd, SADSAD \rightarrow DSADD_{SAD}. Isocentric SAD keeps the source-to-point distance fixed; SSD does not. Scatter factors still differ.

Purpose

Purpose: compute DSADD_{SAD} from DSSDD_{SSD}, SSDSSD, dd, SADSAD in Radiotherapy via DSADDSSD=(SSD+dSAD)2\dfrac{D_{\mathrm{SAD}}}{D_{\mathrm{SSD}}}=\left(\dfrac{\mathrm{SSD}+d}{\mathrm{SAD}}\right)^{2} D_SAD / D_SSD = ((SSD+d)/SAD)². Inverse-square between the two setups. Use it when a real radiotherapy question must be answered in SI before a code check.

Live realistic example

In symbols

Live case. Given DSSD=2.000GyD_{SSD} = 2.000\,\mathrm{Gy}, SSD=100.000cmSSD = 100.000\,\mathrm{cm}, d=10.000cmd = 10.000\,\mathrm{cm}, SAD=100.000cmSAD = 100.000\,\mathrm{cm}, the governing relation DSADDSSD=(SSD+dSAD)2\dfrac{D_{\mathrm{SAD}}}{D_{\mathrm{SSD}}}=\left(\dfrac{\mathrm{SSD}+d}{\mathrm{SAD}}\right)^{2} yields DSAD=2.420GyD_{SAD} = 2.420\,\mathrm{Gy}. A couch, two geometries, two inverse-square spheres. Move a slider: the numbers are this situation, not a canned story.

Calculator

Inputs

Outputs

  • SAD dose D_{SAD}2.420 Gy
Reading speed

Watch on YouTube

Free library

Full library

Free PDF / open book

YouTube channels

RTH-15 · dose
00:0 / 00:08

Narration of this film

A couch, two geometries, two inverse-square spheres.

Isocentric SAD keeps the source-to-point distance fixed; SSD does not. Scatter factors still differ.