RTH-24
Junction dose from gap error
D_j / D0 ≈ 1 + k Δg / P. Gap error times penumbra slope.
Governing equation
where
- \Delta g
- Gap error (mm)
- P
- Penumbra 80/20 (mm)
- k
- Slope factor (—)
- D_j/D_0
- Junction dose ratio (—)
The figure, part by part
Each numbered marker wears the color of its card. Hover a card to light the marker. Tap a marker to jump to that card. The slider walks a red point along the curve; Animate drawing traces it like a pen.
Governing law — red
Domain of the variable — blue
Response curve — green
Rate of change — amber
Lecture brief
Historical brief
The linear-quadratic model, BED, tissue-maximum ratio and Clarkson scatter turned radiotherapy into fraction arithmetic. MU calculation on these sheets is that clinical closed form. This sheet (RTH-24 — Junction dose from gap error) is the form associated with Field matching. Working symbols: , , . A 2 mm gap error on a 6 mm penumbra is a ~15–30 % junction miss. Half-beam blocks shrink this.
Purpose
Live realistic example
In symbols
Calculator
Inputs
Outputs
- Junction dose ratio D_j/D_01.200 —
Watch on YouTube
Free library
Full libraryFree PDF / open book
- Anatomy and Physiology 2eOpenStax · CC BY · Free PDF / open book
- Radiation Oncology Physics (Podgorsak)IAEA STI/PUB/1196 · Free PDF / open book
- IAEA handbook page + teaching slidesIAEA Human Health Campus · Free PDF / open book
- Radiation protection in medical uses of ionizing radiationIAEA · Free PDF / open book
YouTube channels
Narration of this film
Two overlapping penumbras and a hot or cold junction.
A 2 mm gap error on a 6 mm penumbra is a ~15–30 % junction miss. Half-beam blocks shrink this.