NMD-11
Gamma-camera sensitivity
S = C / (A t). Counts per second per MBq in air.
Reading speed
Gamma cameraNEMA NU-1
Governing equation
where
- C
- Counts (—)
- A
- Activity (MBq)
- t
- Live time (s)
- S
- Sensitivity (cps/MBq)
Lecture brief
Historical brief
MIRD schema, in-vivo decay, uptake fractions and PET coincidence turned unsealed sources into organ dose. The sheets compute activity, residence and well-counter geometry. This sheet (NMD-11 — Gamma-camera sensitivity) is the form associated with NEMA NU-1. Working symbols: , , . LEHR collimators run ~70–100 cps/MBq. Sensitivity and resolution trade through hole geometry.
Purpose
Purpose: compute from , , in Nuclear medicine via S = C / (A t). Counts per second per MBq in air. Use it when a real nuclear medicine question must be answered in SI before a code check.
Live realistic example
In symbols
Live case. Given , , , the governing relation yields . A flood or point source, a count total, a sensitivity. Move a slider: the numbers are this situation, not a canned story.
Calculator
Inputs
Outputs
- Sensitivity S208.3 cps/MBq
Reading speed
Watch on YouTube
Free library
Full libraryFree PDF / open book
- Anatomy and Physiology 2eOpenStax · CC BY · Free PDF / open book
- Nuclear Medicine PhysicsIAEA · Free PDF / open book
- Radiation protection in medical uses of ionizing radiationIAEA · Free PDF / open book
- LibreTexts MedicineLibreTexts · CC · Free PDF / open book
YouTube channels
00:0 / 00:08
Narration of this film
A flood or point source, a count total, a sensitivity.
LEHR collimators run ~70–100 cps/MBq. Sensitivity and resolution trade through hole geometry.
Reading speed
Watch on YouTube